|本期目录/Table of Contents|

[1]屠 岩,屠彬彬,朱海清,等.指针压力表智能检定系统误差分析及检测精度评价方法[J].工业仪表与自动化装置,2023,(04):98-102.[doi:10.19950/j.cnki.cn61-1121/th.2023.04.018]
 TU Yan,TU Binbin,ZHU Haiqing,et al.Error analysis of intelligent verification system of pointer pressure gauge and measurement accuracy evaluation method[J].Industrial Instrumentation & Automation,2023,(04):98-102.[doi:10.19950/j.cnki.cn61-1121/th.2023.04.018]
点击复制

指针压力表智能检定系统误差分析及检测精度评价方法

《工业仪表与自动化装置》[ISSN:1000-0682/CN:61-1121/TH]

卷:
期数:
2023年04期
页码:
98-102
栏目:
出版日期:
2023-08-15

文章信息/Info

Title:
Error analysis of intelligent verification system of pointer pressure gauge and measurement accuracy evaluation method
文章编号:
1000-0682(2023)03-0098-05
作者:
屠 岩1屠彬彬2朱海清1黎少鸿1
1.江南大学 机械工程学院, 江苏 无锡 214122;
2.浙江省计量科学研究院, 浙江 杭州 310018
Author(s):
TU Yan1TU Binbin2ZHU Haiqing1LI Shaohong1
1. College of Mechanical Engineering, Jiangnan Universityg, Jiangsu Wuxi 214122, China;
2. Zhejiang Province Institute of Metrology, Zhejiang Hangzhou 310018, China
关键词:
指针压力表智能检定误差分析不确定度检测精度
Keywords:
Pointer pressure gaugeIntelligent verificationerror analysisuncertaintydetection precision
分类号:
TH823.2
DOI:
10.19950/j.cnki.cn61-1121/th.2023.04.018
文献标志码:
A
摘要:
该文首先介绍了指针压力表检定要求与检定流程及智能化检定技术对压力表检定工艺的改进;随后提出了一套指针压力表智能检定系统并介绍了各部分主要功能;接着通过14点稳压标准压力试验分析指针压力表智能检定系统的误差,根据正态性检验、重复性分析及不确定度分析的结果表明,指针压力表智能检定系统造压精度满足0.05%FS要求,且误差呈正态分布。最后提出了指针压力表示值检测精度评价方法,为智能化检定装置提供了分级依据。
Abstract:
This paper first introduces the verification requirements and verification process of pointer pressure gauge and the improvement of intelligent verification technology to the verification process of pressure gauge. Then a set of intelligent verification system of pointer pressure gauge is put forward and the main functions of each part are introduced. Then, the error of the intelligent verification system of pointer pressure gauge was analyzed through the 14-point standard pressure test. According to the results of normality test, repeatability analysis and uncertainty analysis, the pressure building accuracy of the intelligent verification system of pointer pressure gauge meets the requirement of 0.05%FS, and the error is normally distributed. Finally, a method to evaluate the detection accuracy of pointer pressure representation value is proposed, which provides a classification basis for intelligent verification device.

参考文献/References:

[1]伍开宇. 机械指针式压力表的智能检测系统研究[D]. 无锡:江南大学,2022.

[2]渠晴晴. 仪表检定中压力供给系统的设计与开发[D].青岛:中国石油大学(华东),2014.
[3]国家质量监督检验检疫总局.弹性元件式一般压力表、压力真空表和真空表检定规程:JJG 52-2013 [S].2013-12-24.
[4]黄石敏. 自动压力校验系统的设计与开发[D].杭州:中国计量学院,2012.
[5] 国家质量监督检验检疫总局.压力控制器检定规程:JJG 544-2011[S].2012-06-28.
[6]国家质量监督检验检疫总局.测量不确定度评定与表示:JJF 1095.1-2012[S].2013-06-03.
[7]王同宾, 许瑞丰. 精密压力表全省比对参考值示值误差测量不确定度评定[J].工业计量,2022,32(S1):45-47.
[8]蔡一凡. 测量不确定度估算在测量仪器选择中的应用[J].铁道技术监督,2022,50(07):15-18.
[9]丁娟娟, 李新峰. 对压力表测量结果不确定度评定的探讨[J].橡塑技术与装备,2022,48(07):1-5.
[10]粟正宇,杨修能.一般压力表示值误差测量结果的不确定度评定[J].工业计量,2020,30(06):64-65.
[11]陈光辉. 一般压力表的示值误差不确定度的简单评定[J].流体测量与控制,2022,3(01):19-21.

相似文献/References:

备注/Memo

备注/Memo:
收稿日期:2023-02-09

基金项目:
《基于电磁平衡原理的活塞式压力计自动校准系统关键技术研究》项目(2021MK046)

第一作者:
屠岩(1997—),男,江苏淮安人,硕士研究生,研究方向为机械仪表(压力表)智能化检测。

通信作者:
朱海清(1965—),男,江西丰城人,江南大学机械工程学院副教授,硕士生导师,研究方向为为过程装备、特种设备检测技术与检测设备研发。E-mail:809075521@qq.com