|本期目录/Table of Contents|

[1]李 季,章吉刚.气体超声流量测量中的补偿算法及实现[J].工业仪表与自动化装置,2024,(04):110-113+119.[doi:DOI:10.19950/j.cnki.cn61-1121/th.2024.04.021]
 LI Ji,ZHANG Jigang.The compensation in the measurement of gas ultrasonic flow[J].Industrial Instrumentation & Automation,2024,(04):110-113+119.[doi:DOI:10.19950/j.cnki.cn61-1121/th.2024.04.021]
点击复制

气体超声流量测量中的补偿算法及实现(PDF)

《工业仪表与自动化装置》[ISSN:1000-0682/CN:61-1121/TH]

卷:
期数:
2024年04期
页码:
110-113+119
栏目:
出版日期:
2024-08-15

文章信息/Info

Title:
The compensation in the measurement of gas ultrasonic flow
文章编号:
1000-0682(2024)04-0110-04
作者:
李 季章吉刚
(浙江威星智能仪表股份有限公司,浙江 杭州 310015)
Author(s):
LI Ji ZHANG Jigang
(ZheJiang Viewshine High-Tech Co.Ltd.,Zhejiang Hangzhou 311100,China)
关键词:
多声道气体超声流量计补偿算法可靠性对称声道故障声道
Keywords:
multi-channel gas ultrasonic flow-meter compensation algorithm reliability symmetrical channel faulty channel
分类号:
TH814
DOI:
DOI:10.19950/j.cnki.cn61-1121/th.2024.04.021
文献标志码:
A
摘要:
多声道超声流量计相比单声道超声流量计有更好的流场适应性、更高的测量精度和更优的测量可靠性。单声道气体超声流量计出现故障时,流量计不能正常工作,计量功能失效;而多声道超声流量计有冗余性,当其中1个声道失效时仍能保持较高的计量精度,为维修人员争取一定的抢修时间。目前国内很多流量计产品在某些声道损坏的情况下并不能保证计量准确性。该文针对出现故障声道设计了1种基于正常声道数据和各声道独立校表的补偿算法,保证多声道气体超声流量计出现1路故障声道时仍能准确计量。对比测试了不同型号4声道气体超声流量计无故障声道、出现1路故障声道时带补偿算法和不带补偿算法时的示值误差和重复性,以验证补偿算法的有效性。该研究对于设计多声道气体超声流量计的工程人员有一定的指导意义。
Abstract:
Compared with mono channel ultrasonic flowmeter, multi-channel ultrasonic flowmeter has better flow field adaptability, higher measurement accuracy and better measurement reliability. When the mono gas ultrasonic flowmeter fails, the flowmeter cannot work normally and the metering function fails. The multi-channel ultrasonic flowmeter has redundancy, when one of the channels fails, it can still maintain a high measurement accuracy, and strive for a certain rush repair time for maintenance personnel. At present, many domestic flowmeter products cannot guarantee the measurement accuracy when some sound channels are damaged. In this paper, a compensation algorithm based on normal channel data and independent calibration of each channel is designed to ensure that the multi-channel gas ultrasonic flowmeter can still measure accurately when there is one faulty channel. In order to verify the effectiveness of the compensation algorithm, the error and repeatability of different models of 4-channel gas ultrasonic flowmeters with no fault channel, with and without compensation algorithm are compared and tested. The research has a certain guiding significance for the engineers who design multi-channel gas ultrasonic flowmeter.

参考文献/References:

[1]鲍敏.影响气体超声波流量计计量精度的主要因素研究[D].杭州:浙江大学,2004.

[2]李跃忠.多声道超声气体流量计的建模与仿真[J].华中科技大学学报,2006,34(4):39-42.
[3]刘旸.气体超声流量计在声道故障状态下的应用技术研究[J].化工自动化及仪表,2020,47(4):303-307.
[4]全国石油天然气标准化技术委员会.用气体超声流量计测量天然气流量:GB/T 18604—2014[S].2014。
[5]祝飘霞.DN804声道气体超声波流量计设计与适应性研究[D].南昌:东华理工大学,2020.[6]郑丹丹.速度分布对气体超声流量计声传播规律的影响[J].天津大学学报,2017,50(11):1169-1175.
[7]郭燕军.Z型多声道超声波流量计常见故障分析[J].工业计量,2022,32(1):78-80.
[8]李玉军.气体超声波流量计现场应用技术研究[D].西安:西安石油大学,2016.
[9]陈鹏. 超声波流量计诊断及故障处理[J].石化技术,2016(3):127-128.
[10]裴全斌.天然气超声流量计的应用与故障处理[J].石油工业技术监督,2012,28(3):47-48.
[11]全恺.超声流量计现场故障诊断及处理方法[J].计量与测试技术,2009,36(3):33-34.
[12]刘丹丹.多声道超声波气体流量测量若干问题的研究[D].杭州:浙江大学,2017.
[13]王清伟.一种提高超声波流量计测量精度的补偿算法[J].声学与电子工程,2009(3):34-36.

相似文献/References:

备注/Memo

备注/Memo:
收稿日期:2023-12-08第一作者:李季通信作者:章吉刚(1983—),男,浙江嵊州人,硕士,中级工程师,主要从事超声测流技术研究。E-mail:zhang-593@163.com