[1]梁 琦,李 铭,朱凌建,等.基于标准条码线性拟合水平准线偏差的溯源方法[J].工业仪表与自动化装置,2026,(01):81-87.[doi:10.19950/j.cnki.CN61-1121/TH.2026.01.015]
 LIANG Qi,LI Ming,ZHU Lingjian,et al.Traceability method based on standard barcode linear fitting for horizontal alignment deviation[J].Industrial Instrumentation & Automation,2026,(01):81-87.[doi:10.19950/j.cnki.CN61-1121/TH.2026.01.015]
点击复制

基于标准条码线性拟合水平准线偏差的溯源方法()

《工业仪表与自动化装置》[ISSN:1000-0682/CN:61-1121/TH]

卷:
期数:
2026年01期
页码:
81-87
栏目:
出版日期:
2026-02-15

文章信息/Info

Title:
Traceability method based on standard barcode linear fitting for horizontal alignment deviation
文章编号:
1000-0682(2026)01-0081-07
作者:
梁 琦1李 铭1朱凌建2黄秋红2杨晓磊3
1.广西壮族自治区计量检测研究院,广西 南宁 530007;2.西安理工大学,陕西 西安 710048; 3.广西壮族自治区自然资源产品质量检验中心,广西 南宁 530200
Author(s):
LIANG Qi1 LI Ming1 ZHU Lingjian2HUANG Qiuhong2YANG Xiaolei3
1.Guangxi Zhuang Autonomous Region Institute of Metrology & Test, Guangxi Nanning 530007, China; 2.Xi’an University of Technology, Shaanxi Xi’an 710048, China; 3.Guangxi Natural Resources Product Inspection Center, Guangxi Nanning 530200, China
关键词:
检定装置溯源方法水平准线偏差误差分析不确定评定
Keywords:
verification devices traceability method deviation of horizontal alignment error analysis of digital level uncertainty evaluation
分类号:
P224.1
DOI:
10.19950/j.cnki.CN61-1121/TH.2026.01.015
文献标志码:
A
摘要:
基于水平准线偏差的应用原理,针对双目标平行光管数字水准仪检定装置,提出1种互检法的溯源思路。数字水准仪检定装置是1种专用型计量标准装置,可用于光学式水准仪和数字式水准仪的检定。该检定装置主要由2支平行光管组成,光管内部的分划板具有标准条码分划影像。相同的2支平行光管呈对径方向设立作为观测的双目标。该装置按要求建立的水平准线应不大于2″;根据水平准线偏差的误差来源及相关实验数据分析数字水准仪检定装置核心技术指标测量结果的不确定度,结果验证显示该方法建立水平准线的测量结果不确定度优于0.6″,该思路构建的校准方法完全满足量值传递的技术要求,为深入研究数字水准仪检定装置的溯源奠定了良好的基础。
Abstract:
Based on the application principle of horizontal guideline deviation, this paper proposes a traceability approach using mutual inspection method for the verification device of dual-target parallel light pipe digital level. The digital level verification device is a dedicated standard equipment for verification, which can be used for the verification of optical levels and digital levels. The verification device mainly consists of a parallel light pipe with built-in standard barcode grating images. Two identical parallel light pipes are set up diametrically opposite each other as dual targets for observation. The horizontal guideline established by the dual-target parallel light pipes according to requirements must meet the technical requirement of a horizontal guideline deviation of ≤2″. Based on the error sources of horizontal guideline deviation and related experimental data analysis, the uncertainty of measurement results for the core technical indicators of the digital level verification device is determined. The results show that the uncertainty of measurement results for establishing horizontal guidelines using this method is better than 0.6″. The calibration method constructed based on this idea fully meets the technical requirements for value transfer, laying a good foundation for further research on the traceability of digital level verification devices.

参考文献/References:

[1] 全国几何量长度计量技术委员会.水准仪检定装置:JJG 960-2012[S]. 2012.

[2] 全国几何量长度计量技术委员会.水准仪检定规程:JJG 425-2003[S].2003.
[3] 测量不确定度评定与表示:JJF 1059.1-2012 [S].2012.
[4] 广西壮族自治区市场监督管理局.数字水准仪检定装置校准规范:JJF(桂)145-2024[S]. 2024.
[5] 黄秋红,潘莎莎,刘超,等.一种数字水准标尺的新型编码规则及识读方法[J].仪器仪表学报,2023,44(12):244-251.?/div>
[6] 沈迎光,吴学文,董旭明,等.数字水准仪室内快速检校的原理与方法[J].测绘地理信息,2020,45(06):43-46.?/div>
[7] 叶正真,杨维芳,杨锟,等.新型数字水准仪室内检定系统[J].测绘工程,2020,29(03):63-66.
[8] 梁琦,贾真,吴媚,等.数字水准仪室内外检定方法的分析与比较[J].中国计量,2019(06):122+124+123.?/div>
[9] 孙彦锋,周颖熙.数字水准仪测量所需最短条码长度的研究[J].计量与测试技术,2021,48(03):48-50.?div>[10] 沈迎光,吴学文,王露露.专用光管法数字水准仪检校技术研究[J].测绘地理信息,2020,45(05):62-65.
[11] 缪锦根.《数字水准仪检定规程JJG(测绘)2101-2013》中几个问题的商榷[J].北京测绘,2018,32(02):250-254.?/div>
[12] 李亚娟.数字水准仪i角的检校及问题探讨[J].计量技术,2018(01): 61-62.
[13] 赵海军.数字水准仪系统检定关键技术的研究[D]. 西安:西安理工大学, 2017.
[14] 杨俊志,刘宗泉.数字水准仪的测量原理及其检定[M].北京:测绘出版社,2005.
[15] 李文一,程增杰.基于“专用光管法”的数字水准仪检定装置研制[J].中国测试,2017,43(02):73-76.

相似文献/References:

[1]赵泾雄,宋祉霖.嵌入式智能电表检定装置时间激活误差自动控制方法研究[J].工业仪表与自动化装置,2025,(01):88.[doi:10.19950/j.cnki.CN61-1121/TH.2025.01.016]
 ZHAO Jingxiong,SONG Zhilin.Research on automatic control method for time activation error of embedded intelligent electricity meter calibration device[J].Industrial Instrumentation & Automation,2025,(01):88.[doi:10.19950/j.cnki.CN61-1121/TH.2025.01.016]

备注/Memo

备注/Memo:
收稿日期:2025-07-31基金项目:广西壮族自治区市场监管局计量技术规范制修订计划项目[2024-1期-19]第一作者:梁琦(1980—),女,广西南宁人,高级工程师,主要从事测绘仪器计量检定研究。E-mail:hzy09190619@163.com通信作者:李铭(1978—),男,广西南宁人,高级工程师,主要从事大容量容积计量检定研究。