冷水水表动态检定装置研究

《工业仪表与自动化装置》[ISSN:1000-0682/CN:61-1121/TH]

卷:
期数:
2021年04期
页码:
118-121+125
栏目:
出版日期:
2021-08-15

文章信息/Info

Title:
Research on dynamic verification device of cold water meter
作者:
林明星李新民江航成
金卡智能集团股份有限公司,浙江 杭州310018
Author(s):
LIN MingxingLI XinminJIANG Hangcheng
Goldcard Smart group Co.,Ltd.Zhejiang,Hangzhou 310018 ,China
关键词:
水表检定装置活塞式标准器动态流量测试检定用水量流量稳定度
Keywords:
water meter verification device piston type calibrator dynamic flow test verification water consumption flow stability
分类号:
TB937
DOI:
10.19950/j.cnki.cn61-1121/th.2021.04.024
文献标志码:
A
摘要:
冷水水表量大面广,为提高检测效率,提出一种基于主动活塞式的标准器,采用流量时间法的水表动态检定装置。该文介绍了流量时间法测试工作原理和数学计算模型,并对活塞式水表检定装置的关键技术、硬件测控系统、上位机程序进行详细说明。该检定装置体积误差在0.05%以内,标准器重复性为0.02%,装置流量稳定性为0.15%。经试验,测试3台DN15旋翼湿式冷水水表,采用流量时间法用时5分钟左右,而用启停法测试需要时间要42分钟左右,相比常用的启停法提高效率8倍以上,最大误差偏差为0.33%,具有试验流量范围宽、流量准确稳定、工作效率高、能耗低等显著优点。
Abstract:
In order to improve the detection efficiency, a dynamic verification device of water meter based on active piston standard is proposed. Its good stability and repeatability, compared with the start-up and stop method to eliminate the influence of the motion inertia of the detection element, compared with the commutation method to avoid the random error caused by the asymmetry of the flow beam. In this paper, the working principle and mathematical calculation model of flow time method are introduced, and the key technology, hardware measurement and control system and upper computer program of piston water meter verification device are described in detail. The piston type calibrator is a vertical structure with a nominal volume of 40L. The volume error is within 0.05%, the repeatability of the standard is 0.02%, the flow stability of the device is 0.15%, and the expanded uncertainty of the device is better than 0.2% (k = 2). For the water meter with small diameter of DN 15-dn25, the water consumption corresponding to each verification flow point is only 1.0 L, 1.0 L and 20 L. the piston can complete the indication error test of the specified three flow points in one operation. It takes about 5 minutes to test 3 sets of wet type water meters of rotor blades by using flow time method, while it takes about 42 minutes to test with start stop method. The efficiency is improved by more than 8 times, and the maximum error deviation is 0.33%. The device has the advantages of wide test flow range, accurate and stable flow rate, high working efficiency and low energy consumption.

参考文献/References:

[1]全国流量计量技术委员会液体流量分技术委员会.饮用冷水水表:JJG 162-2019[S].北京:中国计量出版社,2020.

[2]孙向东,李斌,马宇峰.静态容积法流量标准装置的构架[J].自动化仪表, 2007,28(4):12-15.
[3]马龙博,郑建英,赵建亮.液体流量标准装置中开式换向器对流量测量误差实验研究[J].自动化仪表, 2015,36(04):60-62+67.
[4]赵建亮,杜伟鹏,马龙博,等.连续测量活塞式液体流量标准装置的设计和应用[J].自动化仪表2016,37(12):32-36.
[5]詹志杰,沈文新,林志良,等.全自动活塞式流量时间法水表检定装置的设计[C].//2012年全国几何量精密测量技术学术交流会.2012:153-158.
[6]沈文新,陆佳颖.双计时法在水表校验装置中的应用探讨[J].计量技术,2008(06):45-46.
[7]全国流量容量计量技术委员会.水表检定装置:JJG 1113-2015[S].北京:中国计量出版社,2015.
[8]王坚.基于三菱PLC的步进电机监控系统的设计与开发[J].北京工商大学学报(自然科学版),2008(05):50-53.

相似文献/References:

备注/Memo

备注/Memo:
收稿日期:2020-12-07

作者简介:
林明星,金卡智能集团股份有限公司检测校准实验室主任,CNAS授权签字人,测试中心副总监,技术评审委员会委员。主要从事实验室管理、计量技术研究工作。
更新日期/Last Update: 1900-01-01